Data Storage; Reset; Resetting The Measurement Settings - wtw Cond 7310 Operating Manual

Conductivity meter
Table of Contents

Advertisement

Cond 7310
ba75933e04
06/2013
Menu item
System /Clock
System /Service infor-
mation
System /Reset

6.2.2 Data storage

This menu contains all functions to display, edit and erase stored measured
values and calibration records (see section 7 D
6.3

Reset

You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
6.3.1

Resetting the measurement settings

The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for conductivity measurements are reset to the default
settings with the Reset function:
Setting
Cal. interval
Measured parameter
Cell constant (C)
(calibrated)
Cell constant (C)
(set)
Temperature compensation
Possible set-
Description
ting
Time
Settings of time and date.
Date
For details, see section 4.4.5
Date format
E
XAMPLE
TING THE DATE AND TIME
Hardware version and software
version of the meter are dis-
played.
-
Resets the system settings to
the default values.
For details, see section 6.3.2
R
ESETTING THE SYSTEM SET
TINGS
ATA STORAGE
Default settings
150 d
ϰ
0.475 cm
(nominal cell constant of the conduc-
tivity measuring cell last calibrated)
0.475 cm
nLF
Settings
2
ON NAVIGATION
, page 20
, page 34
, page 35).
-1
-1
or 0.840 cm
-1
: S
-
ET
-
33