Technical data
Dimensions
Weight
Mechanical structure
Electrical safety
Test certificates
Ambient
conditions
Power
supply
USB interface
Applicable directives
and standards
52

12.2 General data

Cond 7310:
Cond 7310 P:
Cond 7310:
Cond 7310 P:
Type of protection
Protective class
CE, cETLus
Storage
Operation
Admissible relative
humidity
Batteries
Rechargeable batter-
ies
Operational life
Power pack
Type
Baud rate
Data bits
Stop bits
Parity
Handshake
Cable length
EMC
Approx. 240 x 190 x 80 mm
Approx. 290 x 190 x 80 mm
Approx. 0.8 kg
Approx. 1.0 kg
IP 43
III
-25 °C ... +65 °C
+5 °C ... +55 °C
With the power pack connected:
+5 °C ... +40 °C
Yearly mean: < 75 %
30 days/year: 95 %
Other days: 85 %
4 x 1.5 V alkali-manganese batteries, type
AA
4 x 1.2 V NiMH rechargeable batteries,
type AA
(no charging function)
Up to 800 h without / 100 h with illumination
Kuantech Co. Ltd. KSAC 0900110W1UV-1
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 270 mA
Output: 9 V = / 1.1 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
USB 1.1
USB B (device), data output
Adjustable:
1200, 2400, 4800, 9600, 19200 Baud
8
2
None
RTS/CTS
Max. 3 m
EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Cond 7310
ba75933e04
06/2013