Calibration Interval - wtw Cond 7310 Operating Manual

Conductivity meter
Table of Contents

Advertisement

Settings
Setting the calibra-
tion interval
30
Menu item
Measurement /
Temp. comp.
(TC)
/Reference temp.
Measurement
/TDS factor
Measurement /
Stability control
Measurement /
Temperature unit
Reset

6.1.2 Calibration interval

The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
To ensure the high measuring accuracy of the meter, calibrate after
the calibration interval has expired.
The calibration interval is set to 150 days in the factory.
You can change the interval (1 ... 999 days):
1.
Open the menu for measurement settings with [Menu]<F1>.
2.
In the Calibration / Calibration interval menu, set the calibration interval
with <><>.
3.
Confirm the setting with <ENTER>.
4.
Quit the menu with <M>.
Possible
Description
setting
20 °C
Reference temperature
25 °C
This setting is only available for the
measured parameters,
0.40 ... 1.00
Factor for TDS value
On
Switches on or off the automatic stabil-
Off
ity control during measurement (see
section Stability control (AutoRead) &
HOLD function, page 23 )
°C
Temperature unit,
°F
degrees Celsius or degrees Fahren-
heit.
All temperature values are displayed
with the selected unit.
-
Resets all sensor settings to the deliv-
ery condition (see section 6.3.1 R
TING THE MEASUREMENT SETTINGS
page 33).
Cond 7310
ϰ
ρ
and
.
-
ESET
,
ba75933e04
06/2013