Procedures: Current (Source) - GE Druck DPI611 User Manual

Portable pressure calibrator
Hide thumbs Also See for Druck DPI611:
Table of Contents

Advertisement

User Manual
5.6

PROCEDURES: CURRENT (source)

When recalibrating measure and source functions for this range, the
measure function must be adjusted before the source function.
Do the procedure as follows:
1. Connect the applicable calibration equipment
(See Table
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4 ) to do a two-point
calibration (0.2 mA and FS):
4. Check the calibration is correct:
Page 71 of 79
):
5-1
CH1 (24 mA range): (See Table 5-3).
CH1: 24 mA.
Select the applicable Current (source) function
Apply the values that follow:
mA:
Check the error is within limits (See Table 5-3)
0.2, 6, 12, 18, 24
[EN] English – K0571 Issue 1