ADTS542F/552F/553F/554F User Manual
5.1

Introduction

This section details the before-use tasks and the weekly inspection to be carried out by the
operator. The maintenance chart shows the maintenance tasks, the periodicity of each task and a
code referenced to the tasks detailed in Table 5.2.
5.2

Maintenance Tasks

Code
Task
A
Check that all the equipment is present, record any deficiencies.
Visually inspect the exterior of the ADTS and its associated equipment for obvious signs of
damage, dirt and the ingress of moisture. If necessary, use mild liquid detergent and a lint-
free cloth to clean the external surfaces, see the "Safety and Installation" guide K0554.
Inspect the pressure outlet ports for ingress of dirt and moisture, if necessary, clean with a lint-free
cloth.
B
Visually inspect the pneumatic output connectors for damage.
Inspect the small o-ring on each pneumatic output connector for cuts and any signs of wear; replace
as necessary.
Visually inspect pneumatic hoses, electrical cables for cuts, splits and damage; replace as necessary.
C
Before use, power-up the unit as detailed in the "Safety and Installation" guide K0554. Check the
date of the last calibration and, if necessary, refer to the manufacturer.
Record any error messages and refer to Section 6.
D
Daily and before use, carry out the SST and leak check detailed in Section 6.
K0553 Revision C
CHAPTER 5 MAINTENANCE
Task
Code
Inspect
Inspect
Test
Test
Table 5-1 Maintenance Chart
Table 5-2 Maintenance tasks
A
Daily, before use
B
Weekly
C
Before use
D
Daily, before use