ADTS542F/552F/553F/554F User Manual
3. Calculate the actual applied pressure reading to at least 4 decimal places and enter this value into
the ADTS.
4. Repeat this procedure for the other channel.
The adjustment requires two pressures for each channel. Adjustment pressures are the lowest and
highest pressures of the selected channel, see table 4-5:
Following a calibration adjustment use the sensor precision values as the calibration check limits.
Sensor precision value is displayed under the Tools, System status, ADTS, Sensors, Ps/Pt screen.
Sensor precision values are intended for <24hour post adjustment verification of the ADTS against
the calibration equipment used to adjust the ADTS. For >24hour specification values or where
other calibration equipment was used to adjust the calibration refer to the ADTS sales data sheet
accuracy specification.
4.5

Completion of sensor calibration

After completion of all calibration adjustment procedures, carry out the following:
• Make sure the calibration standard and the ADTS are at atmospheric pressure. Disconnect the
calibration standard from the ADTS.
• If there is no further calibration or testing required, exit menus, select standby or switch off the ADTS.
K0553 Revision C
ADTS542F
Ps Channel
Pt Channel
92mbar
92mbar
1130mbar
1997mbar
Table 4-5 Adjustment points
ADTS552F
Ps Channel
Pt Channel
71mbar
71mbar
1130mbar
1997mbar
4-5