Download Print this page

Table Of Contents; Introduction - GE Druck DPI 832 User Manual

Electrical loop calibrator
Hide thumbs Also See for Druck DPI 832:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Table of Contents

Introduction ......................................................................... 1
Safety ..................................................................................... 1
Marks and symbols on the instrument ................................. 2
To start .................................................................................. 2
Location of items ............................................................................ 2
Items on the display ...................................................................... 2
Prepare the instrument ................................................................ 2
Power on or off ................................................................................ 3
Set up the basic operation ......................................................... 3
Select a task (Measure and/or supply) .................................. 3
Set up the settings ......................................................................... 4
Edit functions .................................................................................... 4
Operation .............................................................................. 6
Electrical connections .................................................................. 6
Communications port connections ........................................ 6
Measure mV or Volts ..................................................................... 6
Change the output values .......................................................... 6
Supply mV or Volts ......................................................................... 7
Supply mA .......................................................................................... 7
mA measurements ........................................................................ 7
Switch test .......................................................................................... 8
UPM Pressure measurements .................................................. 8
Error indications .............................................................................. 9
Maintenance ........................................................................ 9
Clean the unit ................................................................................... 9
Replace the batteries .................................................................... 9
Calibration ............................................................................ 9
Before you start ............................................................................... 9
Procedures (mV or Volts input) .............................................. 10
Procedures (mA input) ............................................................... 10
Procedures (mV or Volts output) ........................................... 11
Procedures (mA output) ............................................................ 11
Procedures (IDOS UMM) ............................................................ 11
Specification data ............................................................ 12
General ............................................................................................. 12
Electrical connectors (A1 - Item 10) .................................... 12
Electrical connectors (A2) ......................................................... 12
Customer service ............................................... Back cover
© 2007 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.
K393 Issue 3

Introduction

The DPI 832 Electrical Loop Calibrator is part of the
Druck DPI 800 series of hand held instruments.
The DPI 800 series uses Intelligent Digital Output Sensor
(IDOS) technology to give instant plug and play
functionality with a range of Universal Measurement
Modules (UMM). Example: the Universal Pressure Module
(UPM).
The DPI 832 includes these functions:
Function
* Measure dc mA, mV, Volts
* Supply dc mA, mV, Volts
Step/Ramp functions: Automatic/Manual
Communications port: IDOS or RS232
Language selection
** Measure pressure/Leak test: External IDOS UPM
** Snapshot: Up to 1000 displays with a date/time stamp
®
HART
resistor
Switch test
Other functions: Hold, Maximum/Minimum/Average, Filter, Tare,
Scaled values, Backlight, Alarm
*
Refer to "Specification data".
** Optional item
Safety
Before you use the instrument, make sure that you read
and understand all the related data. This includes: all local
safety procedures, the instructions for the UMM (if
applicable), and this publication.
• It is dangerous to ignore the specified limits for the
instrument or to use the instrument when it is not in
its normal condition. Use the applicable protection
and obey all safety precautions.
• Do not use the instrument in locations with explosive
gas, vapor or dust. There is a risk of an explosion.
• To prevent electrical shocks or damage to the
instrument, do not connect more than 30V between
the terminals, or between the terminals and the
ground (earth).
• UPM only. To prevent a dangerous release of
pressure, isolate and bleed the system before you
disconnect a pressure connection.
WARNING
[EN] English -