3-20
Calibration (Calibrate sensors)
To access this function you will be required to enter your PIN. This function is used to set new
corrected values for the sensors based on the outcome of the calibration procedures detailed in
section 4 "Calibration". The "Calibration" sub-menu contains the following items:
• Sensor: opens the "Calibration check" sub-menu showing the current values for Ps and Pt:
To enter new correction values for Ps:
1. Touch the Ps panel. The Ps sensor correction panel opens.
2. Follow the on-screen instructions. Touch the "tick" icon. A numbered keypad opens.
3. Perform the same procedure to enter new correction values for Pt.
• Change CAL PIN: when selected, this function allows you to enter a new PIN code. To change your
PIN:
1. Touch the "Change CAL PIN" panel. A numbered keypad opens.
2. Enter the new PIN number and touch the "tick" icon. You will be asked to confirm the change.
3. Enter the new PIN again and touch the "tick" icon again. The numbered keypad closes and the
new PIN is now active.
4. Touching the "Cross" icon cancels the action and closes the numbered keypad without
changing the PIN.
Calibration (Software upgrade)
To access this function you will be required to enter your PIN. This function is used to implement
software upgrades for the ADTS Touch and the ADTS following installation of updated software.
The "Calibration" sub-menu contains the following items:
• ADTS Touch software upgrade:
- Application.
- Operating system.
• ADTS software upgrade:
- Main code.
- Boot code.
For a detailed description of the software download and installation procedures see section 5.5,
"Software updates".
Bluetooth
®
Opens a sub-menu containing three items. Functions related to the first two items are only
available when the ADTS Touch is not connected to the mains supply, see also section 3.14
"Bluetooth®":
• List of devices: opens the "List of devices" window containing a list of available devices. Select the
device of interest and touch the "tick" icon. Information related to that device is displayed in the
tools screen. Touching the "cross" icon closes the "List of devices" window.
• New scan for devices: this function initiates a scan of the local area for other ADTS's and ADTS
Touch's, which will then be listed. To select another device:
1. touch on the device name.
2. Touch the "Tick" icon, the device is selected and the list closes.
3. Touching the "Cross" icon, cancels the action and closes the list.