WARNING
OBSERVE SAFETY PRECAUTIONS STATED IN LOCAL ORDERS AND THE AIRCRAFT OR
EQUIPMENT SERVICING PROCEDURES.
C
AUTION
It is the responsibility of the user to make sure that the pneumatic control range
limits are set below the maximum operating limits of the equipment under test.
Make sure the electrical and pneumatic connectors, electrical cables and pipes and positioning of the
ADTS comply with the instructions and requirements in Section 2 Installation.
C
AUTION
D
O NOT USE SHARP OBJECTS ON THE TOUCH SCREEN
DAMAGE THE TOUCH SCREEN
Carry out the following before use:
1. If necessary, carry out the maintenance tasks detailed in Section 5.
2. Make sure the power at the wall connection point is switched OFF. Connect the ADTS to the
electrical power supply at the wall connection point.
MAKE SURE THAT THE SUPPLY INCLUDES A CONNECTION TO A PROTECTIVE EARTH.
Note: Make sure that the power supply switch can be accessed at all times.
3. Inspect the pneumatic hoses for damage, ingress of dirt and moisture. Make sure the aircraft
adaptors are serviceable.
4. Make sure the air vents do not become obstructed.
5. Connect the required hoses to the ADTS that are necessary for the test procedures to be carried
out.
6. Fit the necessary adaptors for aircraft testing to the hoses.
Note: When connected, take care not to kink or stand on the hoses.
7. Fit blanks to all the adaptor test points.
8. Carry out the leak test procedure detailed in Section 6.3.
9. If necessary carry out an altitude correction, see Figure 2-1 also.
Note: Read the whole procedure before starting the test process on an aircraft or component.