- Select the new factor (7) from the list. The new factor is selected and the increment/decrement
factor panel closes.
- Touching the up/down controls (5/6) will increment/decrement the "Rate Aim" value by the
new factor.
5. Touch the "Tick" icon (3), the new value appears in the "Aim" field. The altitude value changes to
the new value, at the rate of change value, and provided that the new value is within acceptable
limits the text will turn green.
6. Touching the "Cross" icon (4), cancels the action and closes the keypad.
• CAS
Displays the currently selected calibrated air speed. To change the value, repeat the same
procedure as above for ALT.
• Mach
Displays the current selection. To change the value, repeat the same procedure as above for ALT.
Ps, Pt and Qc
The following controls are available when "Pressure" units are selected:
• Ps
Displays the currently selected pressure. To change the pressure value:
1. Touch the "Aim" value (1) to highlight it, a numbered keypad is displayed.
2. Use the numbered keypad to input the new value.
3. If required, change the rate of change value by touching the "Rate Aim" value (2) and input the
new value, or use the nudge up/down controls (5/6) to increment/decrement the value by a pre-
defined factor.
4. Touch the "Tick" icon (3), the new value appears in the "Aim" field. The Ps value changes to the
new value, at the rate of change value, and provided that the new value is within acceptable
limits the Ps text will turn green.
K0553 Revision C
Figure 3-6 Nudge increment/decrement factor selection
3-7