Completion

18.
When the calibration check is complete, repeatedly use the QUIT key to return to the first
calibration screen.
i.
If necessary, carry out a rate calibration or press QUIT can be pressed again to exit
calibration mode.
ii.
Select go to ground or stay at the current pressure.
Table 5-1
Calibration Check Points
(both channels)
Calibration Pressure
Point
inHg
1
1
2
5.1
3
10.2
4
15.3
5
20.4
6
25.4
7
30.5
8*
38.7
50.9
10«
63.1
11«
80
* Ps END
« Pt value only
K171 Issue No. 1
mbar
34
173
345
518
691
860
1033
1311
1724
2137
2709
Page 5-5
Ps Calibration Check Points
Point
1
2
3
4
5
6
7
8
9
10
11
Calibration Check Points
Special version
(both channels)
Point
1
2
3
4
5
6*
10«
11«
Table 5-2
Calibration Pressure
inHg
mbar
1
34
3.1
105
5.1
173
7.1
240
10.2
345
15.3
518
20.4
691
25.4
860
30.5
1033
34.6
1172
40
1355
Table 5-3
Calibration Pressure
inHg
mbar
1
34
5.1
173
10.2
345
20.4
691
30.5
1033
38.7
1311
50.9
1724
63.1
2137
77.4
2621
89.6
3034
103.4
3502