Women and athletes are usually thoracic breathers.
Men and obese patients are usually abdominal breathers.
2
If the patient is an abdominal breather, place the respiration belt
around his abdomen.
– or –
If the patient is a thoracic breather, place the respiration belt
around his chest.
As an alternative, the respiratory cushion and belt can be used in
combination with the Body 18.
3
Slide the respiratory cushion underneath the respiratory belt.
The respiratory cushion is attached.
cushion
1
Use the application cushion and position it on the patient together
with the PERU.
2
For respiratory triggering alone, position the electrode leads
loosely on the respiratory belt.
3
Connect the plug for the respiratory cushion to the appropriate
socket on the PERU.
4
Ensure that the pressure hose of the respiration cushion is not
crimped or bent. Make sure the respiration cushion is not unduly
compressed.
With quiet patients, a periodic signal will appear on-screen.
Perform the examination. See:
4.6

External triggering

4.6.1

Input for external trigger signal

The external trigger signal has to meet the following specifications:
Physiological imaging
(
Page 89 Preparing the measurement)