Siemens MAGNETOM Skyra fit Operator's Manual page 68

Mr system
Hide thumbs Also See for MAGNETOM Skyra fit:
Table of Contents

Advertisement

3
MR system components
Troubleshooting: Problems
during docking
68
2
Position the table so that the docking nose is positioned in the
docking station and the table is positioned at the end stop of the
docking station.
The display shows if the table is positioned correctly and
completely in the docking station.
3
Check the display if the table is ready for docking. If the table is not
ready for docking, reposition the table in the docking nose.
4
Press the right pedal down fully with your foot to lock the table to
the MR system - in doubt press the pedal twice.
5
Wait for the table to get ready after docking.
The "Wait" message at the display disappears and the tabletop
moves into the reference position.
Do not undock the table again, before the tabletop is in the
reference position.
Table cannot be docked to the system?
Position the table again so that the docking nose is positioned
correctly and completely in the docking station. Try again to
press the right pedal.
Pedal cannot be pressed
Check that the tabletop is completely in Home position and the
emergency undocking handle is not activated/pulled (reset with
left pedal).
Control units flash
Make sure that the Table Stop buttons are not pressed (at the
table and at the intercom) and that the intercom is powered on.
Table cannot be lowered
Remove any weight from the docking station and the table carriage
cover.