 WARNING
Movement and/or alignment of implants and prostheses in the
body, as well as attraction, alignment, and projectile-like
acceleration of magnetizable objects may result in very serious
hazards!
Injury to patient and operating personnel
Do not use resuscitation devices - for example, defibrillators
or oxygen bottles - in the examination room.
Do not use transport trolleys, movable beds, stretchers, etc.
that consist of magnetizable parts.
Do not wear or carry any magnetizable objects on your person
- for example, watches, pens, scissors, etc..
Use only proven MR Safe or MR Conditional accessories, parts
subject to wear and tear, and disposable articles with the MR
system.
Use only MR Safe or MR Conditional tools and devices.
Service work on the MR system may only be performed by
Siemens Service.
Ensure that only authorized personnel enter the controlled
access area (0.5 mT exclusion zone), for example, electricians
or cleaning personnel trained in MR safety.
Keep the door to the examination room closed.
 WARNING
Magnetizable objects introduced into the magnetic field become
projectiles!
Injury to patient and operating personnel
Inform the operating personnel about the effect of forces on
ferromagnetic parts or implants, especially on systems with a
higher field strength (for example, 3T). Explain that the forces
increase with the field strength of the magnet!
Safety