Physiological imaging
Informing the patient
Attaching the respiratory
cushion and belt
96
4.5

Respiratory triggering

4.5.1

Description

To keep respiratory artifacts to a minimum, respiratory triggering is
primarily used in abdominal imaging. Data acquisition for respiratory
triggering begins when the respiratory signal reaches a predefined
level (approx. 20 % of the maximum value). Respiratory movement is
minimal in this range.
(1) Expiration
(2) Inspiration
The respiratory signal is acquired using a respiratory cushion
connected via a pressure hose to the ECG and respiratory sensor
(PERU). The respiratory cushion is attached to the patient via the
respiratory belt.
Retrospective gating cannot be applied.

Performing

4.5.2
1
Ask the patient to lie still during the measurement.
2
Inform the patient that the knocking sounds during the
measurement are caused by switching the gradients on and off.
The PERU may also vibrate slightly.
The plug of the respiration cushion is NOT connected.
 CAUTION
Incorrect MR image due to disconnected respiratory cushion!
Incorrect diagnosis
As a last step, plug the connector of the respiratory cushion
into the allocated jack.
1
Determine whether the patient is a thoracic or abdominal breather.