Safety
Emergency Shut-down switch
Table Stop button
44
Typically there are two Emergency Shut-down switches installed:
one near the alarmbox and another one in the examination room.
The switch is used to switch off the electric power of the entire MR
system.
 WARNING
Fire or electrical accidents!
Personal injury
Press the Emergency Shut-down switch immediately.
Contact your fire department.
It is located on the intercom and on the patient table.
2.4.2

Medical emergency

 WARNING
Medical emergency during MR measurements!
Risk of death to patients
Terminate the measurement immediately.
Remove patients from the examination room for treatment
unless it is certain that the medical equipment required is
appropriate for use inside an MR room.
Do not store or operate oxygen bottles, defibrillators, or other
auxiliary tools for resuscitation in the examination room.
MR System | Operator Manual
Print No. M6-04401.621.02.02.02