Siemens MAMMOMAT NovationDR Wiring Diagrams
Hide thumbs Also See for MAMMOMAT NovationDR:

Advertisement

MAMMOMAT Novation
Wiring Diagrams
Print No.: SPB7-250.844.01.03.02
Replaces: SPB7-250.844.01.02.02
DR
© Sieme ns AB 2004
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 08.04

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MAMMOMAT NovationDR

  • Page 1 MAMMOMAT Novation Wiring Diagrams © Sieme ns AB 2004 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_...
  • Page 2: Document Revision Level

    Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
  • Page 3: Table Of Contents

    ......2 - 1 MAMMOMAT NovationDR - Back (seen from the right side) ... 2 - 2 LIST OF BOARDS AND FUSES, STAND .
  • Page 4 Changes to previous version ......7 - 1 MAMMOMAT Novation SPB7-250.844.01 Page 4 of 4 Siemens AG Wiring Diagrams Rev. 03 08.04...
  • Page 5: System Overview

    X041E SYSTEM OVERVIEW This page/chapter is intentionally left blank and will be supplied with an update. Siemens AG SPB7-250.844.01 Page 1 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 6 X041E SYSTEM OVERVIEW This page/chapter is intentionally left blank and will be supplied with an update. Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 7: Placement Of Pc-Boards And Components, Stand

    X898, R806 Detector array S862 S896 S895 X858 D750 D803 X801 F801-808 Incoming T802 T801 mains Fig. 1 Front of the stand Siemens AG SPB7-250.844.01 Page 1 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 8: Mammomat Novationdr - Back (Seen From The Right Side)

    Mains input converter S883 S884 T803 X885 X804 Z801 X803 PB0853 Mains input Fig. 2 Back of the stand (seen from the right) Siemens AG SPB7-250.844.01 Page 2 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 9 Stand with rotating support arm Stand with rotating support arm with one FD (Flat Detector) object table and one 18 x 24 object table (used with Opdima) and Mo/W X-ray tube. Siemens AG SPB7-250.844.01 Page 3 of 16 Mammomat Novation Medical Solutions Rev.
  • Page 10: List Of Boards And Fuses, Stand

    400 VAC Brick unit- mounted fuses 2 AT 230 VAC 2 AT 230 VAC NOTE! See spare parts list for replacement fuses. Siemens AG SPB7-250.844.01 Page 4 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 11: List Of Switches And Components, Stand

    3-17/1A Grid out switch located in object table 2. 3-17/2A Grid front switch located in object table 2. 3-17/2A Grid rear switch located in object table 2. Siemens AG SPB7-250.844.01 Page 5 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04...
  • Page 12: Potentiometers

    Transformer for PXCM unit (Brick unit) T803 Transformer for Opdima system Z801 n.a. Filter for Opdima system C801 - C803 n.a. EMC capacitors Siemens AG SPB7-250.844.01 Page 6 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 13: List Of Signals And Test Points, Stand

    X041E LIST OF SIGNALS AND TEST POINTS, STAND This page is intentionally left blank. Siemens AG SPB7-250.844.01 Page 7 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 14 FM OUT 2A D814 3-16/4G FM OUT 2B D814 “ SIGNAL” = “SIGNAL” active low **plus decompression button on control console. Siemens AG SPB7-250.844.01 Page 8 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 15 Filter / Mirror disk step motor M5. FM OUT 2A Filter / Mirror disk step motor M5. FM OUT 2B Filter / Mirror disk step motor M5. Siemens AG SPB7-250.844.01 Page 9 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04...
  • Page 16 D802 D802 PGM_SPEED n.a. POT_RETURN D801 n.a. PRES_FORCE R861 D801 n.a. PRESET_ANGLE R871 D805 D801 “ SIGNAL” = “SIGNAL” active low Siemens AG SPB7-250.844.01 Page 10 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 17 Return line for potentiometers connected to 0VA via a 20Ω resistor. PRES_FORCE Analog signal as preset value for maximum compression force. PRESET_ANGLE Analog signal as preset value for requested rotation angle. Siemens AG SPB7-250.844.01 Page 11 of 16 Mammomat Novation Medical Solutions Rev.
  • Page 18: List Of Voltages And Test Points, Stand

    D805 D801 3-6/4D D801 D750 D801 TXD n.a. ZERO D804 D804.X843 pin 4 “*SIGNAL” = “SIGNAL” active low. ** = Not used Siemens AG SPB7-250.844.01 Page 12 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 19 Indicates that the stereo angle is 0° or ±10°. Visible through vent opening behind tube. TUBE ANGLE Analog signal indicating the rotation angle. Output for sending data via MPS. ZERO Used for adjusting offset of the compression force amplifiers. Siemens AG SPB7-250.844.01 Page 13 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04...
  • Page 20 D805 D802/+14V Digital IC D802/F1 F803 n.a. +5V_R D805 User switches D802/F1 F803 n.a. +12V D805 +14V V_MOT D814 D802/F1 F803 Siemens AG SPB7-250.844.01 Page 14 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 21 Regulated DC voltage, 4.40V < +5V_R < 5.25V +5V_SP Regulated DC voltage, 4.85V < +5V < 5.15V 8.2V_SP Regulated DC voltage, 7.8V < +8.2V < 8.6V Standby typical value Floating voltage. Measure differentially. Siemens AG SPB7-250.844.01 Page 15 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04...
  • Page 22 2-16 X041E LIST OF VOLTAGES AND TEST POINTS, STAND For incoming mains voltage, see page 4-1. NOTE Siemens AG SPB7-250.844.01 Page 16 of 16 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 23: Collimator 1

    COLLIMATOR MAG BOARD DETECTOR ARRAY OBJECT TABLE 1 OBJECT TABLE 2 STEREO OPTION X608 X885 X847 X845 X846 X894 X832 CONTROL CONSOLE X893 X847 X858 X849 X804 X830 X869 X831 FOOT SWITCHES X872 X871 X839 X873 WING BOARD X857 X860 D805 X858 X855...
  • Page 24 INCOMING MAINS P.E. STAND GROUND RAIL GENERATOR GROUND RAIL Door CONTROL PANEL X803 CURTAIN D741 D740 Common Tank UPPER X896 X730 HV Inverter D-subplate X897 TUBE FRONT REAR COVER CHASSIS COVER 0V_sec LIFT TUBE D810 CARRIAGE COMPRESSION PLATE X731 UNIT TUBE X713 HOUSING...
  • Page 28 D801 RESET RESET SW COMP RESET <4.6V D801 TESTPOINTS AND LED’ RESET_HW X812 COLL_ST GRID_SP RESET COMP_SP X813 SPEED D750 FILT_P MASTER STATUSLED’s FOCUS_P ROT_POS COLL_P2 COLL_P1 STAND_REQ_HV_OFF X757 X815 12MHz 0VD2 KVA) PROM X816 CPU & GEN_PREP_RDY (VH) FUNC. GRID_0 GRID_RUNNING DATA AND ADDRESS...
  • Page 37 D801 D802 D805 LEFT SIDE BOARD D805 X865 D807 X812 X822 X807 X851 X851 COLL_ST X813 +5V_R X839 MIRROR X856 FILT_P LAMP LAMP FOCUS_P X807 X865 LAMP ON D807 COLL_P2 COLL_P1 LAMP LAMP X823 X831 SMARTPOINT SMART RIGHT SIDE BOARD INTERFACE 17VACF SMARTPOINT...
  • Page 38 D801 X813 D805 EXT COLL CPU & PLD FUNCTION COLL MOT INT X813 X856 X855 EXTERNAL SMART RX SMART TX SMART TX COLLIMATOR SMARTPOINT INTERFACE S897 FILT MOT INT X839 X855 RESET RESET COLL MOT RX X855 COLL MOT TX COLL MOT RX COLL MOT INT FILT MOT INT...
  • Page 39 OBJECT TABLE 1 X849 OBJECT TABLE 2 X869 D805 CONNECTOR +24VF D801 REAR VIEW CPU & FUNC. GRID_OUT GRID_MOTOR_N X856 X813 TABLE_CONFIG_1 TABLE_CONFIG_3 TABLE_CONFIG_2 X847 X845 TABLE CONFIG TABLE_CONFIG_3 (SEE BELOW) MAG TABLE TABLE_CONFIG_0 TABLE_CONFIG_2 X845 GRID_O CASS_LOADED TABLE_CONFIG_1 MAG TABLE X846 GRID_OUT TABLE_CONFIG_0...
  • Page 40: Mains Voltage Connection

    PE = Protective Earth Note 2. Terminals 1 and 2 are not connected internally. Can be used to secure wires when four or five wire power system is used. Fig. 1 Siemens AG SPB7-250.844.01 Page 1 of 10 MAMMOMAT Novation Medical Solutions Rev.
  • Page 41: Generator: List Of Modules And Fuses

    Type: Ferraz Shawmut - This fuse is used to protect the auxiliary board at unwanted failure, it does not protect internal components and is not intended to be a spare part. PC-boards: Master board D750 Siemens AG SPB7-250.844.01 Page 2 of 10 MAMMOMAT Novation Medical Solutions Rev.
  • Page 42 X041E LEFT BLANK This page is intentionally left blank. Siemens AG SPB7-250.844.01 Page 3 of 10 MAMMOMAT Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 43: Generator: Placement Of Modules, Connectors, Leds And Potentiometers

    -15VA +15VA 28V_sec X700 LEDs X701 S700 X707 LEDs DC_BUS_OK DC_BUS_FAIL Fig. 2 Back of the stand (seen from the left) Siemens AG SPB7-250.844.01 Page 4 of 10 MAMMOMAT Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 44: Control Panel Functions

    Green light indicates that the unit is ready for exposure. This indicator has the same function as the green lights on the stand display and the exposure release buttons. Yellow light and an acoustic signal indicate radiation ON. Siemens AG SPB7-250.844.01 Page 5 of 10...
  • Page 45: D750 Master Board: List Of Switches, Test Points

    Signal from the exposure release buttons at the Control panel to the generator. One of two necessary signals, together with the signal HV_ON, to generate high voltage and thus start the exposure. Siemens AG SPB7-250.844.01 Page 6 of 10 MAMMOMAT Novation Medical Solutions Rev.
  • Page 46 1A/V. FIL_CURR_ACT Analog input signal FIL_CURR_ACT, from the generator, that is proportional to the true filament current. The relation between the filament current and the analog signal is 1A/V. Siemens AG SPB7-250.844.01 Page 7 of 10 MAMMOMAT Novation Medical Solutions Rev.
  • Page 47 The frequency of the signal is equal to the anode rotation frequency. Note, PLD measurement of this signal starts 200 ms after the signal ACC_ANODE goes low, mean value of 10 pulses is calculated. Siemens AG SPB7-250.844.01 Page 8 of 10 MAMMOMAT Novation Medical Solutions Rev.
  • Page 48: Potentiometers

    (high voltage). The relation between the tube voltage and the analog control signal is 5kV/V. HV inverter Logic supply voltage: 0 volt analog Note: 0VA = 0V_sec Siemens AG SPB7-250.844.01 Page 9 of 10 MAMMOMAT Novation Medical Solutions Rev. 03 08.04...
  • Page 49: Leds

    Before making any service/maintenance on the generator modules both LEDs H3 and H4 shall be off, thus making sure the DC-bus voltage not is present. The DC-bus voltage discharges below 60 volt within 3 minutes. Siemens AG SPB7-250.844.01 Page 10 of 10 MAMMOMAT Novation Medical Solutions Rev.
  • Page 50: Generator: Block Diagram

    X041E GENERATOR: BLOCK DIAGRAM INPUT OUTPUT Siemens AG SPB7-250.844.01 Page 1 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 51 X041E GROUND, 0V AND PROTECTION GROUND DISTRIBUTION This page/chapter is intentionally left blank and will be supplied with an update. Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat NovationDR Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 52: Generator: Mains Input & Mains Input Converter

    GENERATOR: X041E MAINS INPUT & MAINS INPUT CONVERTER PRINCIPLE DIAGRAM F807 F808 Siemens AG SPB7-250.844.01 Page 1 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 53 X041E LEFT BLANK This page is intentionally left blank. Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 54: D750 Master Board: Principle Diagram

    X764 0VI_RS232 X772 10V5_AC1 GND_PTR +15VA DSR_OPD_WS GND_PTR GND_OPD_WS RXD_PTR X769 PRINTER/ID-FLASHER TXD_PTR T801 0V_RS232 10V5_AC2 X764 X772 GND_PTR 10V5_AC1 GND_PTR Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 55: Generator - Tube Power Supply: Principle Diagram Tube Filament And Focus Selection

    X041E GENERATOR - TUBE POWER SUPPLY: PRINCIPLE DIAGRAM TUBE FILAMENT AND FOCUS SELECTION Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 56: Generator - Tube Power Supply: Principle Diagram Tube Rotation

    X041E GENERATOR -TUBE POWER SUPPLY: PRINCIPLE DIAGRAM, TUBE ROTATION Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 57: Generator - Hv Inverter Principle Diagram Control Logic

    X041E GENERATOR - HV INVERTER: PRINCIPLE DIAGRAM, CONTROL LOGIC Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 58: Generator - Hv Inverter And Hv Tank: Principle Diagram

    GENERATOR - HV INVERTER AND HV TANK: X041E PRINCIPLE DIAGRAM Siemens AG SPB7-250.844.01 Page 1 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 59 5-10 X041E PRINCIPLE DIAGRAM, CONTROL PANEL This page/chapter is intentionally left blank and will be supplied with an update. Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 60: D750 Master Board: Principle Diagram, Aec

    X041E 5-11 D750 MASTER BOARD: PRINCIPLE DIAGRAM, AEC Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...
  • Page 63: Changes To Previous Version

    New information because of T803 in the block diagram cable con- nections. New information in Generator: Block Diagram, because of T803 and Z801. New information in Generator: Contact Diagram, because of T803 and Z801. Siemens AG SPB7-250.844.01 Page 1 of 2 MAMMOMAT Novation Medical Solutions Rev. 03 08.04...
  • Page 64 X041E Changes to previous version This page intentionally left blank. Siemens AG SPB7-250.844.01 Page 2 of 2 MAMMOMAT Novation Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams...

Table of Contents