Attaching the pulse sensor
Performing the examination
MAGNETOM Skyra
fit
| syngo MR E11
Print No. M6-04401.621.02.02.02
4.4
Pulse triggering
4.4.1
Description
Pulse triggering uses the patient's pulse to trigger the measurement.
A pulse sensor is connected to the patient's toe or finger. The first
pulse wave ("premature pulse wave") is used for triggering. This wave
corresponds to the systolic blood pressure.
A number of special sequences support retrospective gating.
Performing
4.4.2
✓
Suitable finger adapter is attached.
1
Ensure that the cable is not bent.
2
Attach the pulse sensor to a finger or on a toe.
3
Ensure that the pulse sensor is attached properly.
The red LED (fault) at the PPU is flashing rapidly?
No analyzable signal is available.
◆
Ensure that the pulse sensor is attached correctly.
✓
The pulse sensor is attached.
◆
Perform the examination. See:
Physiological imaging
Software operator manual
4