Siemens MAGNETOM Skyra fit Operator's Manual page 42

Mr system
Hide thumbs Also See for MAGNETOM Skyra fit:
Table of Contents

Advertisement

2
Safety
Switch
Emergency Shut-
down
Table Stop
Magnet Stop switch
42
Effect
Electric power of the entire MR system
is switched off, but magnet remains at
field
Motorized table movement is stopped
In case of emergency, the relevant switch should be pressed.
(1) Magnet Stop switch
(2) Emergency Shut-down switch
(3) Table Stop button
The Magnet Stop switch triggers a controlled magnet quench
(shutting down the magnetic field). The MR system is not
disconnected from the power.
There are two different versions of the Magnet Stop switch on the
MR system: as an individual switch or as an integral part of the alarm
box. The switches may also be installed in other places of the MR
system.
Emergency
E.g. in case of fire
In case of accidents or injury due to
table movement
MR System | Operator Manual
Print No. M6-04401.621.02.02.02