4
Physiological imaging
98
Voltage-tim diagram for external triggering of the PMU
Name
U
L
U
H
t
(min.)
L
t
(min.)
H
Input current
Input voltage
Internal contact
External contact
The measurement sequence is triggered by the rising edge of the
external signal.
The external trigger signal can be supplied via the connections in the
magnet cover.
Performing
4.6.2
1
Connect the source of the external trigger signal to the cinch jack
trigger input (magnet cover, left).
2
Perform external triggering. See:
Value
0 V ... 0.8 V
2.5 V ... 5 V
10 ms
10 ms
min. 5 mA
max. ± 5 V
+
-
Software operator manual
MR System | Operator Manual
Print No. M6-04401.621.02.02.02