Board D4, Power Supply, Respiratory And Ecg Triggering - Siemens LITHOSTAR Multiline Manual

Hide thumbs Also See for LITHOSTAR Multiline:
Table of Contents

Advertisement

PC Board Replacement
Supply voltage of the collimator
Supply voltage of the large drives (K1/K2 are energized)
Supply voltage of the small drives (K3 is energized)
Supply voltage of the Com controller (DUEP)

Board D4, power supply, respiratory and ECG triggering

84 37 337 K2013,
Check fuses F1 and F2.
Measure the operating voltage (place the board on the extension)
+ 12 V ± 10%
- 12 V ± 10%
GND
Siemens AG
Medical Solutions
MP
X2.1
Gnd
X2.2
MP
X4.4
Gnd
X4.2
MP
X15
Gnd
X14
MP
X17
Gnd
X16
MP
X2.9
Gnd
X2.6
(X1.d/b/z4)
(X1.d/b/z4)
(X1.d/b/z4)
RXL2-120.841.02.01.02
09.05
+8.5 to +9.5 V regulated
+27 V to +35 V
+28 V to +35 V
+28 V to +35 V
+14 V to +15.5 V regulated
wiring diagram J 1042-44
LED V4
LED V11
Page 27 of 72
CS PS 24
LED V13
LED V14
LED V15
LED V16
The - 5V are not used in the
LITHOSTAR Multiline
.
LITHOSTAR Multiline
27
0