Resetting the Emergency
MAGNETOM Skyra
fit
| syngo MR E11
Print No. M6-04401.621.02.02.02
WARNING
Patient rescue during emergency situations, e.g. quench with
failing quench pipe, fire with strong smoke development,
emergency situation involving patient (e.g. heart attack) and
simultaneous power failure!
Personal injury
◆
After releasing the emergency release, pull the tabletop with
the patient manually out of the magnet.
1
Pull the Emergency release (location is indicated by a label)
outward up to the end stop.
The MR measurement is terminated and the table loses the
referenced position.
2
Turn up the pull-out handle and pull the tabletop out of the
magnet.
3
Rescue the patient.
✓
The error regarding the power supply and/or the motorized drive
release
has been removed.
1
Press the Emergency release back into its original position while
shifting the tabletop slightly in any direction until it audibly locks
into position.
2
To reset the table, simultaneously press the Table Up/Inward
button and the Table Down/Outward button fully.
MR system components
3