Siemens MAGNETOM Skyra fit Operator's Manual page 63

Mr system
Hide thumbs Also See for MAGNETOM Skyra fit:
Table of Contents

Advertisement

Resetting the Emergency
MAGNETOM Skyra
fit
| syngo MR E11
Print No. M6-04401.621.02.02.02
 WARNING
Patient rescue during emergency situations, e.g. quench with
failing quench pipe, fire with strong smoke development,
emergency situation involving patient (e.g. heart attack) and
simultaneous power failure!
Personal injury
After releasing the emergency release, pull the tabletop with
the patient manually out of the magnet.
1
Pull the Emergency release (location is indicated by a label)
outward up to the end stop.
The MR measurement is terminated and the table loses the
referenced position.
2
Turn up the pull-out handle and pull the tabletop out of the
magnet.
3
Rescue the patient.
The error regarding the power supply and/or the motorized drive
release
has been removed.
1
Press the Emergency release back into its original position while
shifting the tabletop slightly in any direction until it audibly locks
into position.
2
To reset the table, simultaneously press the Table Up/Inward
button and the Table Down/Outward button fully.
MR system components
3
63