Cleaning The Camera Lens; Cleaning The Data Carrier; Cleaning The Plugs And Connectors; Cleaning The Patient Table And The Straps - Siemens MAGNETOM Skyra fit Operator's Manual

Mr system
Hide thumbs Also See for MAGNETOM Skyra fit:
Table of Contents

Advertisement

6
Maintenance
118
The LCD monitor/video display are highly sensitive to mechanical
damage.

Cleaning the camera lens

6.1.2
Carefully clean the camera lens with a lint-free cloth and lens
cleaner.
6.1.3

Cleaning the data carrier

1
Clean contaminated data carriers with a clean cloth (cotton or
microfiber).
2
Follow the manufacturer's notes when cleaning CD/DVD data
carriers.
6.1.4

Cleaning the plugs and connectors

1
Dampen a soft cloth with water or a diluted household cleaner
solution. Do not use organic solvents such as alcohol or acetone.
2
Carefully wipe the plugs and connectors with the cloth. Do not
touch the contacts.
6.1.5

Cleaning the patient table and the straps

1
Clean the patient table using a liquid household agent.
2
Wash the straps of the patient table at a temperature of 60°C.
3
Use commercial disinfecting materials. However, do not use
solutions with alcohol or acetone.

Cleaning and disinfecting the receptors

6.1.6
1
Do not use cleaners or disinfectants containing alcohol or ether.
2
Do not use hard or sharp objects (e.g. knives or tweezers) to
remove residue.
3
Clean the receptors with a commercially available cleaner. Follow
manufacturer's instructions.
MR System | Operator Manual