Table 14 Recommended Separation Distance Between Portable And Mobile Rf-Communication Equipment And The X-Ray Detector - Varex Imaging XRpad2 4343 User Manual

Digital x-ray detector
Hide thumbs Also See for XRpad2 4343:
Table of Contents

Advertisement

Table 14
Recommended Separation Distance between Portable and Mobile RF-Communication
Equipment and the X-ray Detector
Recommended Separation Distance between Portable and Mobile RF-Communication Equipment
The X-ray detector is intended for use in the electromagnetic environment specified below. The installer,
X-ray system manufacturer, or user of the X-ray detector should assure that it is used in such an
environment.
Rated Maximum Output
Power of the
Transmitter (W)
0.01
0.1
1
10
100
For a transmitter rated at a maximum output power not listed above, the separation distance can be
estimated using the equation in the corresponding column, where P is the maximum output (power
rating of the transmitter in watt [W]) according to the transmitter manufacture and d is the
recommended separation distance in meter (m).
Note: This guideline may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
44
and the X-ray Detector
150 kHz to 80 MHz
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
www.vareximaging.com
800 MHz to 2.5 GHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
XRpad2 4343
d
2
3 .
P
0.23
0.73
2.3
7.3
23

Hide quick links: