Monthly Inspection; Yearly Inspection - Varex Imaging XRpad2 4343 User Manual

Digital x-ray detector
Hide thumbs Also See for XRpad2 4343:
Table of Contents

Advertisement

XRpad2 4343
11.2

Monthly Inspection

Perform the following inspection at least once a month. If there is a problem, immediately
ask your establishment's safety department to contact your dealer, distributor, or device
manufacturer.
Inspection
Execute the Dark Noise
and X-ray Uniformity tests
(refer to the XRpad2 4343
Digital X-ray Detector
Reference Manual).
Make sure that the XRpad
IPU-2 is clean from dirt or
dust.
Check the manufacturing
date of all battery packs to
ensure that their age is not
five years or older.
11.3

Yearly Inspection

Perform the following inspection at least once a year. If there is any problem, immediately ask your
establishment's safety department to contact your dealer, distributor, or device manufacturer.
Inspection
Execute the Dark Noise,
X-ray Uniformity, Bad
Pixel, and Resolution tests
(refer to the XRpad2 4343
Digital X-ray Detector
Reference Manual).
Result
Date
Date
/
/
Good/Bad Good/Bad Good/Bad If there are changes in
Good/Bad Good/Bad Good/Bad Use the instructions in
Good/Bad Good/Bad Good/Bad Replace the battery pack with
Result
Date
Date
/
/
Good/Bad Good/Bad Good/Bad If there are changes in
www.vareximaging.com
Remedy
Date
/
performance, acquire new
calibration files as described
in the Digital Radiography
Software Manual.
Contact your dealer,
distributor, or device
manufacturer if there is any
problem.
"Section 10.5.1.6, Cleaning
the Battery Pack" on page 22
for cleaning.
a new one and recycle the old
one.
Remedy
Date
/
performance, acquire new
calibration files as described
in the Digital Radiography
Software Manual.
Contact your dealer,
distributor, or device
manufacturer if there is any
problem.
37