Environmental Considerations; Table 6 Environmental Considerations - Varex Imaging XRpad2 4343 User Manual

Digital x-ray detector
Hide thumbs Also See for XRpad2 4343:
Table of Contents

Advertisement

XRpad2 4343
10.2

Environmental Considerations

Storage or use of the X-ray detector and power supply in
WARNING
environmental conditions outside the specification may cause fire,
electrical shock, and unknown hazards, which may result in severe
personal injury, death, or substantial product damage or reduced
product lifetime.
Table 6
includes a list of environmental considerations for the transport, storage, and operation
of the X-ray detector.
Table 6
Environmental Considerations
Environment
b
Ambient Temperature
(14 days)
Relative Humidity (14 days)
Atmospheric Pressure (14 days)
c
Vibration
(EN60068-2-64)
c
Shock
(EN 60068-2-27)
Ingress protection rating
a. In original transport container for 365 days.
b. Temp. Gradient: max 4.5 K/hour.
c. Image quality cannot be guaranteed during shock or vibrations.
Transportation/Storage
–10°C to 55°C/0°C to 55°C
10% to 90%
500 hPa to 1060 hPa
2
3
5 m
/s
(10 Hz to 100 Hz)
2
3
1 m
/s
(100 Hz to 2000 Hz)
25 g (duration 6 ms)
IPx4 rated (protection against splashing water)
www.vareximaging.com
a
Operation
10°C to 35°C
20% to 80%
660 hPa to 1060 hPa
2
3
0.5 m
/s
(10 Hz to 100 Hz)
2
3
0.1 m
/s
(100 Hz to 2000 Hz)
2 g (duration 6 ms)