Ensure the battery charger is powered on.
Orient the battery pack to match the orientation of the battery charger, and insert the battery
pack firmly into the battery charger. Keep the XRpad2 LBP in the battery charger until all the
four charge status LEDs maintains a solid green, indicating a full charge. To remove, lift the
battery out of the battery charger.
Do not drop or hit the battery pack against hard objects as this may
WARNING
cause a risk of damage to the battery pack, which may result in
exposure to the corrosive cell contents, fire, or explosion.
10.5.1.2 Installing the Battery Pack
When installing the battery pack in the X-ray detector in place of the XRpad2 Protective Insert, or
to change a used battery pack, remove the used battery pack before installing the new battery
pack (see
"Section 10.5.1.3, Removing the Battery Pack" on page
There is risk of explosion, personal injury, or damage to the product
WARNING
if the battery pack is replaced by a non-OEM approved component.
To install the battery pack:
Fully support the X-ray detector to prevent it from dropping or slipping.
1
Check to ensure the battery compartment of the X-ray detector is clean, dry, and free of
2
foreign contamination or debris. If cleaning is necessary, see
Battery Pack" on page 22
Check to ensure the battery pack is clean, dry, and free of foreign contamination or debris.
3
If cleaning is necessary, refer to
cleaning instructions.
The battery pack is keyed and can only be inserted into the X-ray detector in one orientation.
4
Align the orientation of the battery pack to match the orientation required on the X-ray
detector.
Insert the charged battery pack into the X-ray detector in the corresponding orientation, and
5
gently press on the end cap until the latches secure the battery pack inside the detector.
Push the Power button on the X-ray detector to power on.
6
Check the battery charge status on the X-ray detector. If the battery charge status shows
7
sufficient battery charge is present, the X-ray detector is ready for use. If the battery charge
status shows lower than desired battery charge level, replace the battery with a charged
battery.
20
for cleaning instructions.
"Section 10.5.1.6, Cleaning the Battery Pack" on page 22