10.7.5
Powering Off the X-ray Detector
The X-ray detector is powered off by holding down one of the following for more than
four seconds:
•
X-ray detector
(Figure 1
•
XRpad IPU-2 push button
•
Extended hand switch push button (wired mode)
10.7.6
Temperature Warning Limits
The XRpad2 4343 detector has two temperature limits that prompts the Digital
Radiography Software and displays a temperature symbol on the detector display informing the
user of the temperature.
•
When the near thermal limit occurs (typically 41°C), immediately finish the exam and then
power down the detector to reduce the thermal load.
•
When the maximum thermal limit occurs (typically at 43°C), the detector will power off
automatically, as soon as the ongoing exam has been completed.
10.7.7
General Workflow for Acquiring an Image
Figure 14
shows the procedure for acquiring an image after starting the Radiography
Imaging Software. Details of the Radiography Imaging Software and the X-ray generator are
described in their corresponding operation manuals.
32
[5]) (wireless and wired mode).
(Figure 10
[1]) (wired mode).
www.vareximaging.com
XRpad2 4343