WARNING
Check Settings
When using your smart device, you should confirm that your volume is turned up,
your phone is not muted, and you do not have headphones plugged in. If your
volume is not turned up, the device is muted, or headphones are plugged in, you
will not hear the sound of any notifications, including important alarms. When you
have headphones connected to your Android
headphones and the speaker. On your Apple, they will sound only in the headphones.
Some notifications are silent during the first visual and vibrate notification and then
make a sound on the second notification. If you don't clear the alert, it repeats at
half volume after 5 minutes and at full volume after 10 minutes.
Your alarm and important alerts sound and display information even when your
volume is low or muted. Specifically, if your smart device is on mute, only these
notifications make a sound:
• Glucose Alarm/Alerts:
• Urgent Low
• Urgent Low Soon
• Low Glucose
• High Glucose
• Rise Rate
• Fall Rate
• No Readings Alert
• System Alerts:
• Calibration Required (after 2-hour sensor warmup, only appears when a
sensor code is not used)
• Calibration Error (only appears when a user enters a calibration; calibration is
not required)
• Sensor Expired
Dexcom G6
®
System User Guide
Chapter 5: Set Up Your Display Devices
, alarm/alerts will sound through the
®