Transmitter Electromagnetic Immunity Specifications
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
Receiver Product Specifications
Part Number
Reading Frequency
Dimensions
TX/RX Frequencies
Bandwidth
Maximum Output Power
Modulation
Dexcom G6
®
System User Guide
Appendix F: Technical Information
Transmitter
IEC 60601 Test
Compliance
Level
Level
± 8 kV Contact
± 8 kV Contact
± 15 kV Air
± 15 kV Air
30 A/m
30 A/m
MT24078
Every 5 minutes
Length: >4.2 inches
Width: >2.5 inches
Thickness: >0.55 inches
2.402 – 2.480 GHz
1.39 MHz
2.4 mW EIRP
Gaussian Frequency-Shift Keying
Electromagnetic
Environment Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Magnetic fields from
common appliances are
not expected to affect the
system.