Table 3-A. Concurrence of G6 CGM Readings and YSI Values by CGM Glucose
Range (Adults; n=159)
YSI (mg/dL)
CGM
Glucose
Range
1
(mg/dL)
< 40
<40
13.5% 56.7% 24.0% 3.8% 1.9%
40- 60
1.2% 67.8% 27.9% 2.7% 0.2% 0.1%
61- 80
0.1% 21.3% 61.4% 16.9% 0.3% 0.1%
81- 120
.
121- 160
.
161- 200
.
201- 250
.
251- 300
.
301- 350
.
351- 400
.
>400
.
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Dexcom G6
®
System User Guide
Appendix F: Technical Information
40-
61-
81-
121-
60
80
120
160
0.4% 13.6% 70.3% 15.1% 0.6% 0.0%
.
0.0% 14.2% 64.3% 20.1% 1.3% 0.0% 0.0%
.
.
0.1% 14.5% 56.7% 26.9% 1.5% 0.2% 0.0%
.
.
.
0.2% 12.1% 59.4% 25.4% 2.9% 0.0%
.
.
.
.
.
.
.
.
.
.
.
.
161-
201-
251-
200
250
300
.
.
.
.
.
.
.
.
.
0.1% 13.7% 59.1% 25.3% 1.9%
.
.
0.2% 22.3% 63.4% 13.7% 0.5% 1,212
.
.
.
0.8% 43.9% 52.5% 2.9%
.
.
.
.
301-
351-
>
Total
350
400
400
.
.
.
104
.
.
.
917
.
.
.
2,275
.
.
.
3,782
.
.
3,026
.
2,597
.
2,869
.
2,268
383
5.9% 76.5% 17.6%
34
298