Receiver
Urgent Low Soon Alert
This alert lets you know you're falling quickly, in fact so quickly that you'll be at or
below 55 mg/dL within 20 minutes, no matter where you are now – even if you're in
your target range. This gives you time to act before you go too low.
What you hear, feel, and see
• Initial alert: Vibrates 6 times and beeps 6 times.
• Until confirmed: Vibrates and beeps 6 times every 5 minutes.
• Updates: 30 minutes later, you get another Urgent Low Soon Alert if you're still
falling so quickly you'll be at or below 55 mg/dL within 20 minutes. The default for
update notifications is 30 minutes, however, you can set them to 15 minutes.
App Notification
Dexcom G6
®
System User Guide
Chapter 10: Alarm and Alerts
138