Number of Readings Provided
The G6 is capable of providing a reading every 5 minutes, or up to 288 readings per
day. For a variety of reasons, the G6 may not display a glucose reading and readings
are "skipped." The percentage of readings you can expect to receive from the G6 over
the sensor life is 98.6%. More than 97% of the sensors captured readings at least
90% of the time. For the G6 with auto-applicator, approximately 99% of the sensors
displayed reading every 5 minutes at least 90% of the time. Table 8 below describes
the reading captured rate by each wear day over the sensor life.
Table 8. Reading Capture Rate by Wear Day (n=374)
Wear Day
1
2
3
4
5
6
7
8
9
10
Dexcom G6
®
System User Guide
Appendix F: Technical Information
Number of Sensors
374
368
364
354
348
338
329
314
303
290
Capture Rate (%)
97.6
98.6
98.7
98.6
98.5
98.5
98.2
97.8
97.0
96.4