Table 5-C. Hyperglycemic Alert and Detection Rate Evaluations (Adults, n=159)
Hyperglycemic
Alert Level
(mg/dL)
120
140
180
200
220
240
300
Dexcom G6
®
System User Guide
Appendix F: Technical Information
Alerts
# of
True
alerts
Alert
(n)
Rate (%)
37,061
97.5
32,148
97.2
23,424
96.6
19,586
96.0
15,689
95.6
12,279
94.6
4,211
85.9
False
# of
Alert
events
Rate (%)
(n)
2.5
12,664
2.8
11,175
3.4
8,455
4.0
7,265
4.4
6,143
5.4
5,007
14.1
2,095
Detections
Correct
Missed
Detection
Detection
Rate (%)
Rate (%)
97.6
2.4
96.8
3.2
95.2
4.8
93.6
6.4
91.2
8.8
88.7
11.3
74.8
25.2