Table 1-B. G6 Accuracy to YSI within CGM Glucose Ranges (Adults; n=159)
CGM
Number
Glucose
of paired
Range
1
CGM-YSI
(mg/dL)
<54
383
54-69
1,537
70-180
9,453
181-250
4,093
>250
3,863
CGM readings are within 40-400 mg/dL, inclusive.
1
Table 1-C. G6 Accuracy to YSI within CGM Glucose Ranges (Pediatrics*; n=165)
CGM
Number
Glucose
of paired
Range
1
CGM-YSI
(mg/dL)
<54
90
54-69
262
70-180
3,144
181-250
1,360
>250
916
* Includes pediatric subjects 6-17 years of age; no YSI measurements were taken for
pediatric subjects 2-5 years of age.
CGM readings are within 40-400 mg/dL, inclusive.
1
Dexcom G6
®
System User Guide
Appendix F: Technical Information
Percent
Percent
Percent
within
within
within
15 mg/dL
20 mg/dL
40 mg/dL
YSI
YSI
YSI
84.3
90.6
89.6
95.1
Percent
Percent
Percent
within
within
within
15 mg/dL
20 mg/dL
40 mg/dL
YSI
YSI
YSI
48.9
62.2
85.5
88.5
Percent
Percent
within
within
15%
20%
YSI
YSI
98.4
99.5
74.1
86.8
80.2
92.1
91.1
97.7
Percent
Percent
within
within
15%
20%
YSI
YSI
85.6
96.6
80.0
90.8
83.4
93.5
89.3
95.9
Percent
Mean
within
MARD
bias
40%
(%)
(mg/dL)
YSI
-6.9
13.8
-0.5
11.5
99.3
-2.8
10.9
99.9
-10.0
9.3
100.0
-3.8
7.1
Percent
Mean
within
MARD
bias
40%
(%)
(mg/dL)
YSI
-20.0
26.0
-5.9
13.3
99.5
-0.3
9.7
99.9
-1.2
8.9
99.9
9.2
7.4