System Electromagnetic Immunity Specifications
Immunity Test
Conducted RF
IEC 61000-4-6
(Receiver only)
Radiated RF
IEC 61000-4-3
Dexcom G6
®
System User Guide
Appendix F: Technical Information
System
IEC 60601 Test
Compliance
Level
Level
3 Vrms
6 Vrms
150 kHz to 80
MHz
10 V/m at 80
MHz to 2700
10 V/m
MHz (AM
Modulation)
Electromagnetic
Environment Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Recommended Separation
Distance
d = 1.2 √P 150 kHz to 80 MHz
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
should be less than
a
the compliance level in each
frequency range
.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
 