Pressure Management - ZOLL AEV Operator's Manual

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

Pressure Management

The following table gives the options and ranges for pressure management.
Parameter Window
Primary
PIP
Value
cm H2O
Secondary
PEEP
Value
PS
EPAP
IPAP
Alarm Limits
High PIP
Low PIP
Measured
Mean Airway Pres-
Value
sure
MAP
Paw Waveform
PIP Context Menu
Breath Trigger
Default : ‐2
(Assisted, 
Spontaneous)
906-0731-01-05 Rev. C
Options /
Range
10 to 80
0 to 30
3 to 30
0 to 60
3 to 30
6 to 60
20 to 100
3 to 35, Off
0 to 99.9
0 to 100
-6 to -0.5
ZOLL Ventilator Operator's Guide
Availability/Notes
Volume Target: Measurement
Pressure Target: Control Set-
ting
PIP values greater than 60 cm
H2O require the operator to
perform a separate
confirmation.
AC Modes (ACV, SIMV, CPAP,
BL Modes
Spontaneous Breaths
(SIMV and CPAP )
Spontaneous Breaths
BL
PEEP cannot be within 5 cm
H
O of the PIP High Limit
2
setting.
Adjustment Increments: .5
Models
All
All
All
All
All
All
All
All
All
4-11