Pop Up Messages - ZOLL AEV Operator's Manual

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

A
LARMS

Pop Up Messages

Alarm Name/Information/Message
Silent/Dark Mode Enabled
Popup triggers when the user attempts to begin operation using Silent/Dark mode. In order to proceed the
user must press the Accept button to begin operation. Note: not all devices are provided with the capability
to operate with Silent/Dark mode, please check with your organization.
Message: Press Accept Key to Enter Silent/Dark Mode Now
Requested Compressor Flow Too Low
Popup triggers when the rate/tidal volume/FIO
ity of the compressor. Resolution involves changing a setting to increase the flow required from the compres-
sor if possible. Note: this condition is only possible with infant setting and for FIO2 <25%.
Mitigation/Info: Reduce FiO2, increase BPM, reduce I Time, or increase Vt
Requested Compressor Flow Too High
Popup triggers when the user attempts to adjust the ventilator so that flow from the compressor is >100 l/
min.
Message: Cannot exceed 100 LPM total flow
Requested O2 Flow Too Low
Popup triggers when the rate/tidal volume/FIO
ity of the O
valve. Resolution involves changing a setting to increase the flow required from the O
2
possible. Note: this condition is only possible with infant setting and for FIO
increase BPM, reduce I Time, or increase Vt
Requested O2 Flow Too High
Popup triggers when the user attempts to adjust the ventilator so that flow from the O
Message: Cannot exceed 100 LPM total flow
Total Requested Flow Too High
Popup triggers when the user attempts to adjust the ventilator so that the combined flow from the compres-
sor and O
valve is >100 l/min.
2
Message: Cannot exceed 100 LPM total flow
Requested Compressor Flow Too High
Popup triggers when the user attempts to adjust the ventilator so that combined flow from the compressor
and O
valve is <2 l/min.
2
Message: Cannot flow less than 2 LPM total flow
Alarm Disable
Popup triggers when the user attempts to adjust to disable an alarm by setting the value to 0 or the maxi-
mum value which would render the alarm essentially off.
Message: Confirmation required -- press accept key to disable alarm
BPM Setting Conflict
Popup triggers when the user attempts to set the BPM to a value that would result in an inspiratory time (I
Time) >3 seconds.
Message: I Time cannot exceed 3 seconds
5-26
combination requires a flow that is less than the flow capabil-
2
combination requires a flow that is less than the flow capabil-
2
www.zoll.com
valve if
2
<25%. Message: Increase FiO2,
2
valve is >100 l/min.
2
906-0731-01-05 Rev. C