This chapter describes how to maintain the ZOLL Ventilator to ensure its optimum working
condition and readiness for immediate use. Specifically, this chapter describes how to
Inspect the device
•
Clean the device
•
Replace intake filters, as necessary
•
Store the device
•
Troubleshoot operational problems
•
In addition to the activities that we describe in this chapter, it is also important to perform
preventative maintenance, replacing worn or defective components, as necessary. Only
ZOLL-trained and certified personnel should perform preventative maintenance using
ZOLL's RCS system.
Inspecting the ZOLL Ventilator
You should perform the following physical inspections of the ZOLL Ventilator on a regular
basis:
Ensure that the ventilator is clean and free of visible damage.
•
Inspect all accessories and connectors for signs of damage or excessive wear. Replace worn
•
or defective items.
Examine high pressure hoses for cracking, discoloration, or disfigurement. Examine end
•
connection fittings for damaged threads and sharp edges. Replace worn or defective hoses
-- DO NOT attempt to repair hoses.
Examine the ventilator circuits for damage or wear including cracking or discoloration. If
•
there are signs of physical degradation or the unit is indicating ventilator circuit problems,
replace the circuit.
Examine the filters and replace them if dirty or clogged.
•
Inspect the external AC /DC adapter, line cords, and DC power cables for wear or damage.
•
Replace if worn or damaged.
906-0731-01-05 Rev. C
ZOLL Ventilator Operator's Guide
Chapter 7
Maintenance
7-1