7. Select Start Up Default Values

When you power on the ventilator, the Start Menu appears, through which you choose the
appropriate parameter default values for the patient. You can select the following patient
parameter defaults:
Adult
Pediatric
Mask CPAP -- Continuous Positive Airway Pressure (CPAP)
Custom -- Values saved in a previous session
Last Settings -- Values set for the patient last using the device before powering down
Note:
Adult Default Parameter Values
The Adult default parameter values are as follows:
Adult Parameter Default Values
Mode
BPM
I:E
VT
PEEP
PIP Limit 35
FIO2
Pediatric Default Parameter Values
The Pediatric default parameter values are as follows:
Pediatric Parameter Start-Up Defaults
Mode
BPM
Ti
PIP
PEEP
PIP Limit 30
FIO2
3-8
ENTILATOR
The ventilator can may be configured to automatically select the Adult parameter
defaults at start up.
AC (V)
12
1:3
450
5
21
SIMV (P)
20
0:6
20
4
21
www.zoll.com
906-0731-01-05 Rev. C