Oxygen Input: High Pressure Gas Supply

The external high pressure gas source connects to the device using the high pressure oxygen
input port.
The device attaches to a regulated supply of 40 to 87 PSIG (280 to 600 kPa). The maximum
flow rate of the oxygen supply is 100 liters per minute. This supply can be from a medical
grade oxygen system or oxygen cylinder (USP).
The OXYGEN IN fitting has a male oxygen Diameter Index Safety System (D.I.S.S.) thread.
Note: If external oxygen is connected, the gas pressure must be at least 41-psig (± 2 psig) when
the device performs Self-Check after you power on the unit.
High Pressure Oxygen Supply Hose
A standard 6 foot oxygen hose is available to make the connection to the high pressure oxygen
source. The hose is has compatible fittings between the device an the source identified for use.
(Also see Chapter 6 "Operating Environments"). Hoses are available from ZOLL, or a suitable
alternative as described below can be used as indicated in the table below.
High Pressure Hoses need to comply with ISO standards
Device Side
Connections
DISS
6 ft (maximum 20ft)
Green or White (as determined by local regulations)
Non-conductive
906-0731-01-05 Rev. C
Hose Attributes
ZOLL Ventilator Operator's Guide
Supply Side Connections
Quick Disconnect, DISS, etc.