Infant/Pediatric, Single-Limb, Wye Patient Circuits; Intended Use; Specifications - ZOLL AEV Operator's Manual

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

P
C
ATIENT
IRCUITS

Infant/Pediatric, Single-Limb, Wye Patient Circuits

ZOLL provides the following infant/pediatric patient circuits for use with all models of the
ZOLL defibrillator:
Pediatric/Adult, 6 ft
Pediatric/Adult, 12 ft

Intended Use

The ZOLL Infant/Pediatric, Single-Limb, Wye Patient Circuits (6 and 12 ft.) are intended for
use with all models of the ZOLL Ventilator. ZOLL Ventilators have a single-limb circuit
interface and no internal exhalation valve. Patient circuits are used as a means by which to
transfer breathing gases from a ventilator to a patient (inhalation) and from a patient to
atmosphere (exhalation).
This patient circuit is intended for use when delivering tidal volume from 50 ml to 300 ml.

Specifications

Infant/Pediatric, 6 ft Patient Circuit
The Infant/Pediatric, 6 ft patient circuit (
Internal Diameter: 10 mm
RINSP @ 15 Lpm: 0.11 hPa/l/min
REXP @ 15 Lpm: 0.17 hPa/l/min
C
@ 60 hPa: 0.5 ml/hPa
T
Deadspace: 4.2 ml
Maximum Working Pressure: 100 hPa (cm H
Complies with ISO 5367. Breathing Tube intended for use with Anesthetic Apparatus and
Ventilators.
Non-Sterile - For Single Patient Use Only
Not intended for use with heated humidification.
-40°C to 70°C (-40°F to 158°F)
Caution
During use circuit may come into contact with biohazard material. Handle carefully to avoid
cross-contamination.
Contents of Infant/Pediatric, 6 ft patient circuit package:
72" (1.83 m) Ventilator Wye Breathing Circuit, with cut-to-size tube and exhalation valve
Exhalation drive line (Clear) with connector
Pressure line (Green) with connector
D-8
(REF 820-0107-XX)
(REF 820-0131-XX)
REF 820-0107-XX
www.zoll.com
) has the following specifications:
O)
2
906-0731-01-05 Rev. C