ZOLL AEV Operator's Manual page 153

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

Infant/Pediatric, 12 ft Patient Circuit
The Infant/Pediatric, 12 ft patient circuit (
Internal Diameter: 10 mm
R
INSP
R
EXP
C
@ 60 hPa: 0.8 ml/hPa
T
Deadspace: 4.2 ml
Maximum Working Pressure: 100 hPa (cm H
* The extended length of the tubing in the 12 ft circuit results in a higher R
the 6 ft circuit.
Complies with ISO 5367. Breathing Tube intended for use with Anesthetic Apparatus and
Ventilators.
Non-Sterile - For Single Patient Use Only
Not intended for use with heated humidification.
-40°C to 70°C (-40°F to 158°F)
Caution
During use circuit may come into contact with biohazard material. Handle carefully to avoid
cross-contamination.
Contents of Infant/Pediatric, 12 ft patient circuit package:
144" (3.65 m) Ventilator Wye Breathing Circuit, with cut-to-size tube and exhalation valve
Exhalation drive line (Clear) with connector
Pressure line (Green) with connector
Warning!
Compressible volume can significantly decrease the delivered minute ventilation of
small adults and pediatric patients. When managing patients at risk, always correct for
compressible volume following Step 6 in the following "Directions for Use" section of
this appendix.
Warning!
Do not use this circuit with PEEP settings below 5 cm H
Warning!
Given the additional length of the 12ft circuit, the system may not be able to trap PEEP
in patients with short expiatory time. Always ensure that the device is performing as
required.
906-0731-01-05 Rev. C
@ 15 Lpm: 0.17 hPa/l/min*
@ 15 Lpm: 0.17 hPa/l/min
ZOLL Ventilator Operator's Guide