BPM Parameter Settings

The following table gives the options and ranges for the BPM parameters:
Parameter Window
Primary Value
BPM
Breaths per
minute
Secondary
Ti (sec)
parameter Val-
|or
ues
I:E
Alarm Limits
High breath rate 20 to 99, off
Low breath rate
Measured Value Minute Volume
Vmin (ml)
BPM Context Menu
Control
Default I:E
Parameter
Rise Time -
Default #
Cycle Off %
Default 25%
(% Cycle )
Spont Ti Limit
Default
Adult = 3.00
Infant = 2.00
Mask CPAP =
3.00
906-0731-01-05 Rev. C
Options /
Range
0 to 80
Ti 0.1 to 3.0
or
1:1 to 1:99
Ti 0.1 to 5.0
or
I:E 4:1 to 1:99
2 to 40
0 to 99.9
1:1 to
1:99, 0.3 to 3
Ti 0.1 to 5.0
or
I:E $:1 to !:99
1 to 10
10 to 70%
0.30 to 4.00
ZOLL Ventilator Operator's Guide
Availability/Notes
Volume Target: Control Setting
Pressure Target: Measured
See BPM context menu: Con-
trol Parameter
Inverse I:E
The control value is shown in
the Parameter window, the
dependent value is shown in the
Auxiliary Box.
Inverse I:E
Auxiliary Box
Auxiliary Box
Models
All
All
EMV+,
Eagle II
All
EMV+,
Eagle II
All
All
4-7