M
AINTENANCE
Caution
There are no user serviceable parts except the filter components above.
Caution
When used in dusty/dirty environments, you should check the foam and disk filters, and replace
them as needed. This prevents particle build up on the transducer screen and the need to take the
unit out of service for maintenance by a biomedical technician.
Caution
If filters are exposed to biological matter, dispose of them following Universal Precaution
procedures for your facility.
Note:

Replacing the ZOLL Ventilator's Filters

Tools needed:
Hemostat or tweezers
Phillips Head screwdriver
Warning!
Before attempting to replace filters, make sure that external power is disconnected and
that the ventilator's power switch is set to "OFF".

Replacing the Foam Filter

The Foam Filter is located inside the
Compressor Inlet Fitting.
1. Carefully remove the Foam Filter using
a hemostat or tweezers.
DO NOT reuse or attempt to clean the old filter.
Replace the Foam Filter with a new filter. Lightly tap
2.
the new filter into place. The top of the filter should
reside approximately 3/4 to 7/8" below the height of the
22 mm female connector that is part of the Compressor
Inlet Fitting.
7-4
Do not attempt to clean this filter and do not operate the internal compressor without a
filter in place.
www.zoll.com
906-0731-01-05 Rev. C