Appendix A Specifications; General - ZOLL AEV Operator's Manual

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

General

Parameter
Operating Modes
Breath Target
Flow Rate
Breath Rate
Inspiratory Time (Ti)
Tidal Volume
FIO
2
PEEP/EPAP
Peak Inspiratory Pressure
(PIP)
906-0731-01-05 Rev. C
Operating Range
®
EMV+
and Eagle II
: AC,SIMV, CPAP with and without Pressure Support,
BL Modes, and Leak Compensation for Active and Noninvasive ventilation.
®
AEV
: AC, CPAP (with and without Pressure Support), BL Modes, and Leak
Compensation for Noninvasive ventilation.)
Volume or Pressure
0 to 100 LPM at 40 cm H
O
2
1 to 80 BPM ±1 BPM over the interval
Setting Resolution: 1 BPM
Measurement: 1 to 90 BPM ± 1 BPM over the interval
Setting: 0 to 3 ± s for I:E ratio > 1:1.0, 0 to 5 ± s for E:I ratio <= 1.0:1
Setting Resolution: 0.05 s
Setting: 50 to 2000 ml ATPD ± (5 ml +10% setting)
Setting Resolution: 10 ml
Measurement: 0 to 9999 ml ATPD ± (5 ml +10% setting)
21 to 100% (± 3% of full scale ± 10% of setting)
Setting: 0 to 30 cm H
O ± (2 cm H
2
Setting Resolution: 10 ml
Measurement: 0 to 30 cm H
Setting: 10 to 80 cm H
O (± 2 cm H
2
Setting Resolution: 1 cm H
Measurement: 0 to 99 cm H
ZOLL Ventilator Operator's Guide
Appendix A
Specifications
O + 8% of reading)
2
O ± (2 cm H
O + 8% of reading)
2
2
O ± 8% of setting)
2
O
2
O ± (2 cm H
O + 8% of reading)
2
2
A-1