A
LARMS
PS Conflict
Popup triggers when the user attempts to set the PS > the PIP High limit - PEEP pressure.
Message: Cannot adjust PS higher than PIP High Limit - PEEP
Leak Comp.
Popup triggers when the user attempts to initiate leak compensation (LC). To do this, the user must press
the Accept button and then select LC followed by Accept again to confirm the setting change.
Message: Some Alarms Disabled! Configure Alarms for Patient!
Mode Conflict
Popup triggers when the user attempts to initiate leak compensation (LC) during volume targeted ventila-
tion. Note: LC is only available during pressure targeted ventilation.
Message: Cannot select Volume targeted control breaths with Leak Compensation on -- first turn Leak Com-
pensation off
Inverse I:E
Popup triggers when the user attempts to set an inverse I:E ratio. To do this, the user must press the
Accept button and then adjust the I:E ratio to the desired inverse value and press Accept again to confirm
the setting change.
Message: Confirmation required -- press accept key to allow inverse I:E
Excessive Volume for Infant Circuit
Popup triggers when the user attempts to set a Vt >300 ml when the tubing compliance compensation is set
to Pediatric.
Message: Press accept to confirm use of adult circuit
Insufficient Volume for Adult/Ped Circuit
Popup triggers when the user attempts to set a Vt <200 ml when the tubing compliance compensation is set
to Adult.
Message: Press accept to confirm use of infant circuit
High PEEP Setting
Popup triggers during CPAP mode when the user attempts to set the PEEP >15 cm H
user must press the Accept button and then adjust PEEP to the desired value and press Accept again to
confirm the setting change.
Message: Confirmation required -- press accept key to allow PEEP above 15
High EPAP Setting
Popup triggers during BL mode when the user attempts to set the EPAP >15 cm H
must press the Accept button and then adjust PEEP to the desired value and press Accept again to confirm
the setting change.
Message: Confirmation required -- press accept key to allow EPAP above 15
EPAP Setting Conflict
Popup triggers during BL mode when the user attempts to set the EPAP <3 cm H
Message: Cannot adjust EPAP target to within 3 of IPAP target
EPAP Setting Conflict
Popup triggers during BL mode when the user attempts to set the EPAP <5 cm H
Backup PIP target.
Message: Cannot adjust EPAP target to within 5 of backup PIP
PIP Limit Conflict
Popup triggers during BL mode when the user attempts to set the peak inspiratory pressure (PIP) < the
IPAP target.
Message: Cannot adjust high limit lower than IPAP target
5-30
www.zoll.com
O. To do this, the
2
O. To do this, the user
2
O below the IPAP target.
2
O below the Apnea
2
906-0731-01-05 Rev. C