Pediatric/Adult, Single-Limb, Wye Patient Circuits; Intended Use -- Pediatric/Adult Patient Circuits; Specifications - ZOLL AEV Operator's Manual

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

P
C
ATIENT
IRCUITS

Pediatric/Adult, Single-Limb, Wye Patient Circuits

ZOLL provides the following pediatric/adult patient circuits for use with all models of the
ZOLL defibrillator:
Pediatric/Adult, 6 ft
Pediatric/Adult, 12 ft

Intended Use -- Pediatric/Adult Patient Circuits

The ZOLL Pediatric/Adult, Single-Limb, Wye Patient Circuits (6 and 12 ft.) are intended for
use with all models of the ZOLL Ventilator. ZOLL Ventilators have a single-limb circuit
interface and no internal exhalation valve. Patient circuits are used as a means by which to
transfer breathing gases from a ventilator to a patient (inhalation) and from a patient to
atmosphere (exhalation).
The pediatric/adult patient circuits are intended for use when delivering tidal volume from
200 ml to Adult.

Specifications

Pediatric/Adult, 6 ft Patient Circuit
The Pediatric/Adult, 6 ft patient circuit (
Internal Diameter: 22 mm
R
INSP
R
EXP
C
@ 60 hPa: 2.8 ml/hPa
T
Deadspace: 22 ml
Maximum Working Pressure: 100 hPa (cm H
Complies with ISO 5367. Breathing Tube intended for use with Anesthetic Apparatus and
Ventilators.
Non-Sterile - For Single Patient Use Only
Not intended for use with heated humidification.
-40°C to 70°C (-40°F to 158°F)
Caution
During use circuit may come into contact with biohazard material. Handle carefully to avoid
cross-contamination.
D-2
(REF 820-0106-XX)
(REF 820-0130-XX)
@ 30 Lpm: 0.02 hPa/l/min
@ 30 Lpm: 0.10 hPa/l/min
www.zoll.com
) has the following specifications:
REF 820-0106-XX
O)
2
906-0731-01-05 Rev. C

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eagle iiEmv+

Table of Contents