Oxygen Input: High Pressure Gas Supply; High Pressure Oxygen Supply Hose - ZOLL Z Vent Operator's Manual

Table of Contents

Advertisement

Product Overview

Oxygen Input: High Pressure Gas Supply

An external high pressure gas source connects to the ventilator using the high pressure oxygen
input port. The device attaches to a regulated medical grade (USP) O
supply of 40 to 87 psig (280 to 600 kPa). Maximum flow rate of the oxygen supply is 100 liters
per minute. The Oxygen Input fitting (See Figure 2-9) has a male oxygen Diameter Index
Safety System (DISS) thread.
Note:

High Pressure Oxygen Supply Hose

A standard 6 foot oxygen hose is available for connecting the ventilator to a high pressure
oxygen source. (Also see Chapter 6 "Operating Environments"). Hoses are available from
ZOLL, or a suitable alternative as described below can be used as indicated.
High Pressure Oxygen Hose for compliance with ISO standard (ISO STANDARD 5359)
Ventilator Side
Connections
DISS
2-10
Figure 2-8 Pneumatic Design
If external oxygen is connected, the oxygen pressure must be at least 41 psig (± 2 psig)
at the time the ventilator performs its Self-Check after turning on the ventilator.
Hose Attributes
6 ft (maximum 20 ft) Green or White (as
determined by local regulations)
non-conductive
www.zoll.com
system or O
cylinder
2
2
Supply Side Connections
Quick Disconnect, DISS, etc.
9650-002360-01 Rev. A