PECIFICATIONS
Parameter
Pressure Support (PS)/
IPAP
Oxygen Input Pressure
Mean Airway Pressure
(MAP)
Airway Pressure High
Limit
Airway Pressure Low
Limit
Breath Trigger
Airway Pressure
Waveform
Minute Volume
LED Status/Alarm
Indicator
Alarm Volume
Noise Level
Operating Voltages
Operating Time
Internal Battery
Ventilator Temperature
Ranges
Battery Temperature
Ranges
Size
Weight
Warranty
A-2
Operating Range
0 to 60 cm H
O ± (2 cm H
2
NOTE: IPAP must always be greater than or equal to 3 cm H
55 psig
Reading: 0 to 99.9 cm H
O ± (2 cm H
2
Resolution: 1 cm H
O
2
Setting: 20 to 100 cm H
O
2
Setting Resolution: 1 cm H
Setting: Off, 3 to 35 cm H
2
Setting Resolution: 1 cm H
-6.0 to -0.5 cm H
O (± 0.25 cm H
2
0 to 99 cm H
O ± (2 cm H
2
0 to 99.9 lpm ± 0.1 lpm + 8% of reading
Red, Yellow, and Green
82 dBA @ 1 meter
~60 dBA when measured at 1 meter (operating at default settings using the
compressor only)
AC Supply: 100 to 240 VAC (50/60 and 400 Hz) -- use only the AC/DC power supply
that ZOLL provides with the device.
DC Supply: Nominal 12.5 to 28.0 VDC (accepts DC voltages between 11.8 to 30
VDC).
10 hours at default settings
Operating Extreme: -25
C to 49
Battery Charge: 0
C to 45
8.0" Wide X 12.5" High X 4.5" Deep (20.3 cm Wide X 31.8 cm High X 11.4 cm Deep)
~9.7 lbs (4.4 kg)
Limited, 1 year
www.zoll.com
O + 8% of setting)
2
O +8% of reading)
2
O
2
O
O
2
O + 5% of setting) below
2
O + 8% of reading)
2
C (-18
F to 120
F)
C (32
F to 113
F)
O.